This Hach DR 5000 Spectrophotometer was removed from an operating environment due to a major upgrade. the item show signs of use consistent with its age.
It was tested with the following notes from our testing technician:
- Tested to power on, due to transportation and handling the Self Check Test stops while starting the instrument Showing the following error (“Please close the lid”)..
- the buttons, controls, LED’s and display are in good working order.
- The alarm system and beep work as untended.
- USB Ports are in good working order.
- Filter test pass.
- no further test done (due lid issue)
returns are not accepted, for any further testing please prepare instruction and contact us to get it done for you before you buy and thank you.
As with all scientific, measuring, or precision instrument, this device may need calibration or specialized maintenance prior to use.
Type : LPG408.99.00012
SN: 1248931
For service Password / code : please inquire
Includes:
- 1x Hach DR 5000 Spectrophotometer – For Part
Does Not Include:
- Anything else not listed or pictured above